EasyManua.ls Logo

Siemens MAMMOMAT Inspiration - Test 5: Phantom Imaging Quality; Objective of Phantom Imaging Quality Test; Required Equipment for Phantom Imaging Quality; Procedure for Phantom Imaging Quality

Siemens MAMMOMAT Inspiration
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test 5: Phantom Imaging Quality
MAMMOMAT Inspiration 33
SPB7-330.640.01.01.02
Test 5: Phantom Imaging Quality
Objective
To ensure that adequate phantom imaging quality is achieved
Required Equipment
a) ACR (RMI 156) phantom
b) Compression plate 24 x 30 cm
2
Procedure
1. Create a new patient record in the local database.
Fill in: Last Name: Test_Five
First Name: Phantom Imaging Quality
Patient ID: day+time when the test is performed
(example: 200707071030)
DOB: 05 05 1955
Gender: Other
2. Select procedure QC-raw.
3. Select exposure technique AEC
28 kV
target/filter combination Mo/Mo
grid in
auto decompression off
OPCOMP off
AEC segmentation on
low dose mode off
Change all projection views to L-CC