Test 5: Phantom Imaging Quality
MAMMOMAT Inspiration 33
SPB7-330.640.01.01.02
Test 5: Phantom Imaging Quality
Objective
To ensure that adequate phantom imaging quality is achieved
Required Equipment
a) ACR (RMI 156) phantom
b) Compression plate 24 x 30 cm
2
Procedure
1. Create a new patient record in the local database.
Fill in: Last Name: Test_Five
First Name: Phantom Imaging Quality
Patient ID: day+time when the test is performed
(example: 200707071030)
DOB: 05 05 1955
Gender: Other
2. Select procedure QC-raw.
3. Select exposure technique AEC
28 kV
target/filter combination Mo/Mo
grid in
auto decompression off
OPCOMP off
AEC segmentation on
low dose mode off
Change all projection views to L-CC