Test 5: Phantom Imaging Quality
34 Quality Control Manual
SPB7-330.640.01.01.02
4. Position phantom on object table according to Figure 6.
Figure 6 Position of Phantom
5. Mount compression plate and perform compression.
6. Perform an exposure.
7. Examine the image. For evaluation use magnification, zooming or gray scale
inversion, adapt windowing values, if helpful. If the viewing conditions are
considered to be inferior while looking at the image on AWS, send image to
viewing station for additional examination of the achieved image.
Detector
ACR (RMI 156) phantom
Chest wall side