Test 6: Spatial Resolution (optional)
MAMMOMAT Inspiration 37
SPB7-330.640.01.01.02
Test 6: Spatial Resolution (optional)
Objective
To detect assumed spatial resolution problems in different parts of the detector
Required Equipment
a) Bar pattern phantom (containing line groups with spatial frequencies of 5 lp/mm,
6 lp/mm, 7 lp/mm, 8 lp/mm) applicable for mammography
b) Collimator mountable PMMA phantom
c) Compression plate 24 x 30 cm
2
Procedure
1. Create a new patient record in the local database.
Fill in: Last Name: Test_Six
First Name: Spatial Resolution
Patient ID: day+time when the test is performed
(example: 200707071030)
DOB: 05 05 1955
Gender: Other
2. Select procedure QC-raw.
3. Select:
exposure technique MANUAL
28 kV
90 mAs
target/filter combination clinically used
grid in
auto decompression off
OPCOMP off
4. Mount PMMA phantom close to collimator.
5. Move bar pattern phantom on object table to a position where a problem with the
spatial resolution of the detector is assumed.
6. Mount compression plate and perform compression.
7. Perform exposure.