Test 2: Artifact Detection
MAMMOMAT Inspiration 21
SPB7-330.640.01.01.02
10. If artifacts are still visible on any image after step 9 call Siemens customer
service.
11. Note result in Test form Test 2.
12. Close examination.
Performance Criteria and Corrective Action
No clinically relevant artifacts shall be visible on image.