Test 1: AEC Test
MAMMOMAT Inspiration 15
SPB7-330.640.01.01.02
Test 1: AEC Test
Objective
To test the AEC security cut-off
To measure AEC stability for different object thicknesses
Required Equipment
a) Shielding plate in object table size, e.g. 2 mm steel or lead plate
b) PMMA plates of 10 x 10 cm² and 20 mm, 30 mm, 40 mm, 50 mm, 60 mm,
70 mm thicknesses
c) Compression plate 24 x 30 cm
2
Procedure
1. Create a new patient record in the local database.
Fill in: Last Name: Test_One
First Name: AEC
Patient ID: day+time when the test is performed
(example: 200707071030)
DOB: 05 05 1955
Gender: Other
2. Choose procedure QC-raw.
AEC security (step 3 - 10)
3. Select exposure technique AEC
28 kV
target/filter combination mostly used
grid in
auto decompression off
OPCOMP off
AEC segmentation on
low dose mode off
4. Place shielding plate on object table.
5. Mount compression plate and perform compression.
6. Perform an exposure.
7. Check that the radiation signal lights up very shortly and exposure is interrupted,
tube load is below 5 mAs.
8. Check that a warning is shown.
9. Note result in Test form Test 1.
10. Deactivate examination, remove shielding plate.