POLYMOBIL III RXB8-115.034.02 Page 2 of 2 Siemens AG
Rev. 03 04.05 CS PS 24 Medical Solutions
2 - 2 Unpacking
• Remove the protective covers on the single tank and the collimator.
• Unscrew the 4 hexagonal head screws (2/Fig. 2, width across flats 17 mm) remove the
bow (3/Fig. 2) and the U-rail (4/Fig. 2).
• Bring the foot pedal from the locked position (5/Fig. 2) into the maneuvering position
(6/Fig. 3). For further foot pedal positions refer to the operator manual.
• Remove the 2 hexagonal head screws and the bow (7/Fig. 2, width across flats 17 mm)
on the left and right of the stand carriage.
• Push the POLYMOBIL III over the rail (8/Fig. 3) and move it over the ramp (9/Fig. 3)
away from the pallet.
Fig. 2 Fig. 3