The values that were determined are to be logged and evaluated in the report “Protective
conductor resistance / report” by indicating the measuring points.
In addition, the measuring procedure and the measuring device used (designation and
serial number) are to be documented.
NOTE
The report “Protective conductor resistance / report” is filed in the
“Certificates” register in the system folder.
NOTE
If documentation of protective conductor resistance measure-
ments is missing, a new report must be created.
A new report “Protective conductor resistance / report” is included
in the SPR8-125.034.02.02 startup instructions. Detach this report
from these instructions, complete it and file it in the “Certificates”
register in the system folder. For a first determination, the deter-
mined values are to be logged as values measured for the first
time. Evaluation is not required in this case.