38 Inspection and Maintenance
POLYMOBIL Plus SPR8-125.831.01.02.02 Siemens AG
05.05 CS PS 24
Page 38 of 40
Medical Solutions
2.13 DAP measuring system (optional)
Press the test button on the display. The value on the display should be in the range
80-120 µGym
2
(DAP resolution 0.1 µGym
2
) or 8-12 µGym
2
(when using a chamber of
high sensitivity, i.e., DAP resolution 0.01 µGym
2
). If not, calibrate the system according
to the Service Instructions. (when using a chamber of high sensitivity, i.e., DAP resolu-
tion 0.01 µGym
2
). If not, calibrate the system according to the Service Instructions.
X
Check the function of the DAP measuring system by releasing an exposure. Check if a
measured value is shown on the display.
SIE Function of DAP measuring system