2.12 Device leakage current measurement
SIE Device leakage current measurement
Observe the information on the device leakage current measurement contained in these
instructions.
The device leakage current measurement is to be carried out according to
ARTD-002.731.17... with the unit closed.
For this, the measuring method indicated in the report is to be used. If the value first
measured needs to be determined again, either direct measurement or differential
measurement can be selected.
The limit value of 2.5 mA must not be exceeded.
The measuring method (differential measurement or direct measurement) and the
measuring device used (designation and serial number) are to be documented.
The actual mains voltage is to be measured and logged. If the measured mains voltage
deviates from the nominal voltage, the measuring value is to be corrected to the value
which corresponds to a measurement with the nominal value of the mains voltage. This
is to be additionally documented.
The highest value is to be entered into the report “Device leakage current / report” and
evaluated.
NOTE
The report “Device leakage current / report” is filed in the “Certifi-
cates” register in the system folder.
NOTE
If documentation of device leakage current measurements is miss-
ing, a new report must be created.
A new report “Device leakage current / report” is included in the
SPR8-125.034.02.02 startup instructions. Detach this report from
these instructions, complete it and file it in the “Certificates” reg-
ister in the system folder. For a first determination, the determined
values are to be logged as values measured for the first time. Eval-
uation is not required in this case.

Other manuals for Siemens POLYMOBIL Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens POLYMOBIL Plus and is the answer not in the manual?

Siemens POLYMOBIL Plus Specifications

General IconGeneral
BrandSiemens
ModelPOLYMOBIL Plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals