36 Inspection and Maintenance
POLYMOBIL Plus SPR8-125.831.01.02.02 Siemens AG
05.05 CS PS 24
Page 36 of 40
Medical Solutions
2.11 Protective conductor test
SIE Protective conductor test
Observe the information on the protective conductor resistance test contained in these
instructions.
• The protective conductor test is to be carried out according to ARTD-002.731.17... with
the unit closed.
The protective conductor resistance must not exceed 0.2 Ohm.
• The determined values are to be logged and evaluated in the report “Protective conduc-
tor resistance / report” by indicating the measuring points.
• In addition, the measuring method and the measuring device used (designation and se-
rial number) are to be documented.
NOTE
The report “Protective conductor resistance / report” is filed in the
“Certificates” register in the system folder.
NOTE
If documentation of protective conductor resistance measure-
ments is missing, a new report must be created.
A new report “Protective conductor resistance / report” is included
in the SPR8-125.034.02.02 startup instructions. Detach this report
from these instructions, complete it and file it in the “Certificates”
register in the system folder. For a first determination, the deter-
mined values are to be logged as values measured for the first
time. Evaluation is not required in this case.