Siemens AG SPR8-125.831.01.02.02 POLYMOBIL Plus
05.05 CS PS 24
General 19
Page 19 of 40
Medical Solutions
2.5 Required operator documents com-
plete, available, readable
2.1.3 Customer documents, test:
SI customer documents
Ask the customer for additional required
operator documents and check for com-
pleteness, availability and readability.
2.6 All warnings available and visible? 2.1.2 Labels, test:
SIM labels
2.7 Operating icons, LEDs and key label-
ing ok?
2.9 Radiation, test:
SIE Radiation Indicator
2.10 Control console, tests:
SIE Operating elements
SIE Indicators
3. Electrical checks n.a.
3.1 Protective conductor resistance mea-
surement
2.11 Protective conductor test, test:
SIE Protective conductor test
3.2 Device leakage current or equivalent
device leakage current measurement
n.a.
3.2.1 Device leakage current measure-
ment
2.12 Device leakage current measurement,
test:
SIE Device leakage current measurement
3.2.2 Equivalent device leakage current
measurement
n.a.
3.3 Patient leakage current or equivalent
patient leakage current measurement
n.a.
3.3.1 Patient leakage current measure-
ment
n.a.
3.3.2 Equivalent patient leakage current
measurement
n.a.
4. Mechanical checks n.a.
4.1 Wall, ceiling and floor attachments
fixed and without damage?
n.a.
Test certificate “Technical safety
checks”
Maintenance instructions, test / remarks