Siemens AG SPR8-125.831.01.02.02 POLYMOBIL Plus
05.05 CS PS 24
General 21
Page 21 of 40
Medical Solutions
5.7 Table and unit positions reproducible
(e.g., zero positions, layer height etc.)?
n.a.
6. Product-specific tests 2.2 Checking the screws, tests:
SIM Cassette compartment
SIM Pedals
2.3 Checking the handles, tests:
SIM Single-tank yoke
SIM Single tank
SIM Transport handle above the control
console
2.5 Checking the pedal positions, tests:
SIM Maneuvering movement
SIM Forward movement
2.6 Single tank, test:
SIM Suspension
2.8 Stand, test:
SIM Stand mounting
2.13 DAP measuring system (optional), test:
SIE Function of DAP measuring system
7. Test result / evaluation: Evaluate the long-term tendencies of the
protective conductor resistance and the
device leakage current by comparing the
actual measuring values with the values
determined during previous safety-relevant
tests. A sudden increase of the measured
values, even if the limit values are not
exceeded, indicates a safety-relevant fault.
Enter the result of the test.
Have the TSC protocol signed by the opera-
tor or a person authorized by the operator.
Hand over the TSC protocol to the operator
or a person authorized by the operator.
Test certificate “Technical safety
checks”
Maintenance instructions, test / remarks