SIREMOBIL Compact Register 3 SPR2-130.033.02 Page 4 of 10 Siemens AG
System Manual Rev. 02 12.98 TD SD 24 Medical Engineering
3 - 4 Installing the lifting column
Installing the cover with the control panel 3
• Place the cover over the lifting column.
• Route the ribbon cable and the protective conductor for the control panel through the
opening (1/Fig.9).
• Put the cover on.
• Secure the cover to the lifting column (10) with 4 screws and from below the control lever
side with 4 screws as well.
Two long screws are provided on the inside.
• Reinsert the pin in the control lever.
• Reinstall the strain relief (2/Fig.9) of the control panel cable.
• Reattach (3/Fig.9) the protective conductor to the chassis (on the control side from below
left).
• Reinstall the cover plate (on the control panel side at the chassis, under the control lever)
Fig. 9 Fig. 10
Fig. 11 Fig. 12