Place the new collar over the lifting column, center it and attach it securely.
Place the C-arm back into the mounting (3 people required).
Reinstall the swivel brake:
- Insert the bolt from below through the latches.
- Reinstall the spring pin with its opening facing you.
Install the spring washers:
the lower spring with the cone upward; the upper spring with the cone downward.
Secure the latch assembly with the screws.
Position the brake lever.
Secure the cover plate with the screws under the swivel brake.
Reattach the hand railing to the C-arm with the screws.