3 - 1
Siemens AG Register 3 SPR2-130.033.02 Page 1 of 10 SIREMOBIL Compact
Medical Engineering Rev. 02 12.98 TD SD 24 System Manual
Preliminary steps 3
Recut the(1/Fig.1) threads.
Eject all the wires from the four-pole connector on the connection cable and connect
them to the three-pole connector as follows:
Remove the black plastic cover from the lifting column (1/Fig.2).
Installing the new lifting column 3
Move the new lifting column (two people required) through the opening in the chassis.
Be careful not to damage the connection cable and ensure that the EMC gasket is
seated correctly in the mounting.
Place the lifting column into the guide (1/Fig.1) on the chassis and secure it with the
screws from underneath the unit (screw torque = 40 Nm).
Fig. 1 Fig. 2
X7A/1 X7A/2 X7A/3
1
1
Installing the lifting column 3