SIREMOBIL Compact Register 3 SPR2-130.033.02 Page 6 of 10 Siemens AG
System Manual Rev. 02 12.98 TD SD 24 Medical Engineering
3 - 6 Installing the lifting column
Installing the line filter 3
• Insert the 4 standoffs (1/Fig.13) with the short threads into the holes in the lifting column
and tighten them.
• Insert the base plate and tighten it with 4 M4 nuts and contact washers.
• Pull the cable back behind the lifting column toward the right.
• Secure all protective conductors (2/Fig.14) again on the base plate of the line filter;
when doing this, insert contact washers between the cable terminal and the base plate.
• Secure the line filter with two screws to the base plate.
• Secure the BNC plug to the base plate with the screws.
• Connect the BNC cable.
Installing the power pack M 14 3
• Install the base plate.
• Install the power pack with the plug connections facing down.
Fig. 13
Fig. 14 Fig. 15