Siemens AG Register 3 SPR2-130.033.02 Page 1 of 2 SIREMOBIL Compact
Medical Engineering Rev. 02 12.98 TD SD 24 System Manual
Safety information 1
Tests and adjustments performed with radiation on are identified with the radiation warn-
ing symbol . Radiation protection must be worn during these types of adjustments.
Protective conductor resistance test 1
After concluding all work and reattaching all covers, perform the protective conductor test
according to TI 236 / ARTD-002.731.17... .
The protective conductor resistance may not exceed 0.2 ohm.
Other information 1
n.a.
Required documents 1
Circuit diagram G5429
Service instructions RXR2-130.061.01....
Required tools and measurement devices 1
Required material 1
When performing service work and tests, adhere to the product-specific
safety information contained in the documents as well as the general
safety information in Register 2 of the TI folder.
Service kit (refer to ARTD / Part 3)
Digital multimeter
20N spring scale 44 29 114 NH029
200N spring scale 44 15 113 RH090
25 - 130 Nm 1/2" torque wrench 34 24 561
Service PC
Service-PC - Host connection cable 99 00 440 RE999
Protective conductor meter 44 15 899 RV090
Viscogen KL300, 40 g 73 95 353
Conductive paste for heat sink
Lifting column parts kit 37 79 407 G5429
CAUTION
General 1