Installing the lifting column 3 - 9
Siemens AG Register 3 SPR2-130.033.02 Page 9 of 10 SIREMOBIL Compact
Medical Engineering Rev. 02 12.98 TD SD 24 System Manual
3
Adjusting the swivel brake 3
• Move the C-arm into the horizontal position (Fig.19).
• Measure the braking force with a (20N) spring scale for the released brake and a (200N)
spring scale for the applied brake from the handle grip end to the image intensifier.
(Fig.19);
When doing this, suspend the spring scale with a cloth band or other similar device from
the handle. Set the brake to the above values by adjusting the grip latch on the brake
lever .
• Place the brake lever in the end positions (brake applied or released) respectively,
parallel to the horizontal arm.
• Set the brake.
• Remove the brake lever.
• Glue the plastic guard in the center with a little silicon adhesive.
• Install the brake lever in the same position as previously and secure it with the screws.
Remaining work 3
• Attach the new ring cover with adhesive to the control lever at the center.
• Switch the system on and check all functions.
• Move the lifting column all the way up and attach the scale label supplied vertically in the
middle of the inner section.
• Install all the covers.
When doing this, attach the protective conductors to the covers.
Perform the protective conductor test.
Set point: Brake released e 10 Nm Brake applied E 100 Nm