2 - 1
Siemens AG Register 3 SPR2-130.033.02 Page 1 of 6 SIREMOBIL Compact
Medical Engineering Rev. 02 12.98 TD SD 24 System Manual
Preliminary steps 2
Switch the SIREMOBIL off and disconnect the line voltage plug.
Secure the unit so that it cannot be inadvertently switched on.
To secure the unit against tipping, support it on the operating panel side with wooden
boards or other similar material.
Board D3 2
Remove the Plexiglas cover of board D3 by removing the screws.
Disconnect all plugs except D3/X2 (Fig.1).
Remove the protective conductor from the board.
Board D1 2
Disconnect all plugs from the board.
Remove the protective conductor from the board.
Remove all 6 mounting nuts (1/Fig.2).
Carefully remove the board and place it on an ESD surface.
Remove the spacers from the 6 screws.
Open the strain relief for ribbon cable X 4 (2/Fig.2).
Fig. 1 Fig. 2
Board D3 does not have to be removed.
1
2
NOTICE
Removing the lifting column 2