Removing the lifting column 2 - 3
Siemens AG Register 3 SPR2-130.033.02 Page 3 of 6 SIREMOBIL Compact
Medical Engineering Rev. 02 12.98 TD SD 24 System Manual
Board D2 2
• Remove the protective conductor from the interface plate.
• Remove the interface plate.
• Remove the grounding bracket from the heat sink. The heat sink remains on the board
(1/Fig.3).
• Disconnect all plugs from the board.
• Remove the 6 spacer screws.
• Unscrew the board from the heat sink (1/Fig.4).
• Remove the board and place it on an ESD surface.
• Remove the grounding strap from the heat sink (5/Fig.5).
• Unscrew the protective conductor.
• Remove the 6 mounting screws of the heat sink.
• Lift out the heatsink.
Line filter 2
• Disconnect the lower connections of the line filter.
• Remove the line filter by removing the screws.
• Remove all protective conductors from the base plate of the line filter (2/Fig.5).
• Remove the BNC connector (X1) and remove the screws from the BNC plug (3/Fig.5).
• Unscrew the grounding strap from the base plate (4/Fig.5).
• Remove the screws from the base plate and remove the base plate; remove the bolt from
the lifting column.
Power pack M 14 2
• Disconnect plugs X2 and X3 from the power pack; the protective conductor can remain
connected.
• Remove the power pack;
to do this, loosen only the two upper nuts and remove the lower nut.
TV power supply M13 2
• Remove the Plexiglas cover by removing the screws.
• Desolder the connections from the transformer and disconnect all other connections.
• Remove the screws from underneath the TV power supply and lift it out, Fig.6
• Remove the stabilization bracket of the monitor cable connector from the lifting column
by removing the screws , 1/Fig.7.