Siemens AG SPR2-230.814.03.02.02 SIREMOBIL Iso-C 3D
11.05 CS SD 24
Checks Prior to Installation 15
Page 15 of 52
Medical Solutions
Testing the 3D software 0
Systems with imaging system PC: CELSIUS C650 part no. 71 29 013 0
• After system start-up, press the key combination "Ctrl-D" on the MEMOSKOP key-
board to switch over to 3D operation.
¹ syngo screen appears on right-hand monitor
• Click "Help"
• Click “About ISO-C 3D...”
• Check to see whether the software “VB00D” appears under “3DSW”.
• If the 3D software “VB00D” is installed, continue with the "function control" item
• If the 3D software “VB00D” is not installed, install it according to the instructions found
in the document “Software Description SPR2-230.815.01.xx.xx”.
• Close the window
Systems with imaging system PC: CELSIUS C444 part no. 77 14 145 0
• After system start-up, press the key combination "Ctrl-D" on the MEMOSKOP key-
board to switch over to 3D operation.
¹ syngo screen appears on right-hand monitor
• Click "Help"
• Check to see whether the software “VC00B” appears under “3DSW”.
• If the 3D software “VC00B” is installed, continue with the "function control" item
• If the 3D software “VC00B” is not installed, install it according to the instructions found
in the document “Software Description SPR2-230.815.01.xx.xx”.
• Close the window