36 BrainLAB navigation system
SIREMOBIL Iso-C 3D SPR2-230.814.03.02.02 Siemens AG
11.05 CS SD 24
Page 36 of 52
Medical Solutions
Image quality test 0
Preparation 0
Move the C-arm into the neutral horizontal position.
- The image intensifier is on the main unit.
Take the required stand out of the "Calibration Phantom Service Case".
Attach the resolution test (lead strip test pattern) to the stand so that it is horizontal to
the floor.
Position the resolution test in the isocenter of the C-arm. Navigate from the side and
longitudinal to the C-arm.
Use the laser light localizer (if one is installed) to aid in locating the isocenter.
Fig. 22: Test_set-up_side_view
Fig. 23: Test_set-up_top_view