46 BrainLAB navigation system
SIREMOBIL Iso-C 3D SPR2-230.814.03.02.02 Siemens AG
11.05 CS SD 24
Page 46 of 52
Medical Solutions
Checking 3D high resolution 0
Target:
Determine the optimal slice plane for representing the lead strip test in the syngo window.
In upper left-hand window:
• Using the vertical bar, select the plane perpendicular to the lead strip test pattern.
• Next, select a slice plane in the center of the lead strip test pattern.
Fig. 34: Slice_plane_setting_in_upper_left_window
• If the lead strip test pattern is slanted in relation to the floor plane, correct this by rotat-
ing the slice plane in Free Mode.