Siemens AG SPR2-230.814.03.02.02 SIREMOBIL Iso-C 3D
11.05 CS SD 24
License and configuration 25
Page 25 of 52
Medical Solutions
NOTE
If no other IP address is displayed, then the system has not yet
been assigned its own IP address, and the following IP address
should be assigned to it.
If the system has already been assigned an IP address, then con-
tinue with the item "Navigation DICOM nodes".
NOTE
A system must never be configured with two IP addresses,
since this can lead to software conflicts.
• Select "Specify an IP address".
NOTE
The system cannot be used with a DHCP server.
• Enter 192.17.3.110 as the IP address.
• Enter 255.255.255.0 as the subnet mask.
• Do not select "Enable DNS for Windows resolution".
• Select "Enable LMHOSTS lookup".
WINS
• Do not enter anything under "Primary WINS server".
• Do not enter anything under "Secondary WINS server".
• Do not enter anything under "Gateways".
DNS
• Do not enter anything under "Domain".
• Do not enter anything under "DNS service search order".
• Click "Save".
• Close the service software.
• Switch the system completely off and then back on.
- This ensures that the configuration is completely saved within syngo.