34 BrainLAB navigation system
SIREMOBIL Iso-C 3D SPR2-230.814.03.02.02 Siemens AG
11.05 CS SD 24
Page 34 of 52
Medical Solutions
Set C-arm position as given in setting no. 1 = "angulation 0 degrees and orbital -95 de-
grees."
- Take note of the C-arm positions displayed.
Select "50" as the number of images, as given in setting no. 1.
Click the "Go" button.
Press the left footswitch and hold it down until "Calibration successful" is displayed.
¹ The system automatically starts a measurement run.
Click "OK" after "Calibration successful" is displayed.
Perform this procedure for each of the settings given in the table below.
Tab. 1
Perform these settings suc-
cessively
Setting Number of
images
No. 1 Angulation = 0 degrees
Orbital = -95 degrees
50
No. 2 Angulation = 0 degrees
Orbital = +95 degrees
50
No. 3 Angulation = 0 degrees
Orbital = -95 degrees
100
No. 4 Angulation = 0 degrees
Orbital = +95 degrees
100
Intermediate step At setting no. 5, recheck the
adjustment of the calibration
phantom.
n.a.
No. 5 Angulation = 180 degrees
Orbital = +95 degrees
50
No. 6 Angulation = 180 degrees
Orbital = -95 degrees
50
No. 7 Angulation = 180 degrees
Orbital = +95 degrees
100
No. 8 Angulation = 180 degrees
Orbital = -95 degrees
100