Installing/Mounting
4.4 Mounting
SITRANS F M MAG 5100 W
Operating Instructions, 11/2010, A5E03063678-02
19
4.4 Mounting
Install the sensor in rigid pipelines in order to support the weight of the meter.
Center the connecting pipelines axially in order to avoid turbulent flow profiles.
Use proper gaskets according to liner type
Figure 4-7 Correct installation with gaskets
Vibrations
Avoid strong vibrations.
Figure 4-8 Avoid vibrations
CAUTION
In applications with strong vibrations, Siemens recommend remote mounting of the
transmitter!