Installing/Mounting
4.5 Potential equalization
SITRANS F M MAG 5100 W
Operating Instructions, 11/2010, A5E03063678-02
21
Class 150 AWWA DN PN 10 PN 16 PN 40
800 32" 274/274 202/202 386/380 285/285 N/A N/A N/A N/A 260 192
900 36" 288/288 213/213 408/408 301/301 N/A N/A N/A N/A 240 177
1000 40" 382/382 282/282 546/546 403/403 N/A N/A N/A N/A 280 207
1050 42" N/A N/A N/A N/A N/A N/A N/A N/A 280 207
1100 44" N/A N/A N/A N/A N/A N/A N/A N/A 290 214
1200 48" 395/395 292/292 731/731 539/538 N/A N/A N/A N/A 310 229
1400 54" -/503 -/317 -/736 -/543 N/A N/A N/A N/A 528 389
1600 66" -/684 -/505 -/913 -/674 N/A N/A N/A N/A 698 515
1800 72" -/771 -/569 -/937 -/692 N/A N/A N/A N/A 700 516
2000 78" -/867 -/640 -/1128 -/832 N/A N/A N/A N/A 890 656
4.5 Potential equalization
To obtain optimum results from the measuring system, the sensor must have the same
electrical potential as the liquid being measured.
This is achieved by means of built-in grounding electrodes.
Figure 4-10 Potential equalization with earthing electrodes
Cathodic protected piping
Special attention must be paid to systems with cathodic protection
WARNING
Use in hazardous area!
Cathodic pipe protection is not allowed in hazardous areas