Loading...
Installing/Mounting
4.5 MAG 5000/6000 CT
SITRANS F M MAG 5000/6000
Operating Instructions, 12/2013, A5E02338368-002 29
4.5.1 Installing hardware key
Use hardware key on non-verified transmitter
1. Mount hardware key on transmitter connection plate during setting of primary operating
parameters such as Q
max
., low-flow cut-off, units, approvals, etc. in connection with
commissioning or calibration. See setup menus in appendix menu diagrams.
2. Remove hardware key after setting up and calibrating unit.
This locks the menu structure and the selected settings.
Note
Hardware key function
Setting of primary operating parameters is blocked during normal operation.
When key is mounted, access to all menu items is gained. When key is removed, primary
settings are blocked in accordance with requirements in authorisation.
4.5.2 Seal device
Seal transmitter
CAUTION
Seal transmitter to prevent unauthorized access.