Installing/Mounting
4.6 Turning transmitter/keypad
SITRANS F M MAG 5000/6000
Operating Instructions, 12/2013, A5E02338368-002 31
4.5.3 Installation conditions
4.5.3.1 MI-001
MAG 5000/6000 CT together with MAG 5100W (7ME652) are approved for Mi-001 under the
following installation conditions.
DN 50 to 300 mm (2" to 12")
Horizontal installation
Compact or remote with max. 3 m cable
Power supply 115/230 V AC
Other restrictions may apply (see certificate).
4.5.3.2 PTB K7.2
MAG 5000/6000 CT together with MAG 5100W (7ME652) are approved for PTB K7.2 under
the following installation conditions.
SITRANS F M MAG 5100 W with MAG 5000/6000CT
DN 50 to 300 mm (2" to 12")
Horizontal installation
Compact or remote with max. 10 m (33 ft.) cable
Other restrictions may apply (see certificate 22.76/10.02)
**insert installation conditions as pasted
4.6 Turning transmitter/keypad
It is possible to alter the standard assembly, e.g. to turn transmitter or keypad.