PROFIsafe
1.8 Quit PROFIsafe commissioning
SITRANS P, DS III PROFIsafe series
Supplement to Operating Instructions, 03/2006, A5E00732533-01
1-11
Note
SIMATIC PDM operation
The write protection in the PROFIsafe commissioning status "S4" allows the maintenance
timer to be changed, for example for the calibrating interval of the electronics and for the
servicing interval of the sensor.
Note
If you need a user-defined PIN write protection, you will find it in the appropriate chapter.
See also
Overview (Page 1-6)
Create user-defined PIN and lock the device (Page 1-7)
1.8 Quit PROFIsafe commissioning
1.8.1 Preparations for maintenance and service
Procedure
Before you perform maintenance on a PROFIsafe device, proceed as follows:
1. Deactivate "PROFIsafe Commissioning".
2. Deactivate the PIN write protection if necessary.
1.8.2 Deactivate PROFIsafe commissioning in SIMATIC PDM
Procedure
1. Click on "Device > PROFIsafe Commissioning > Quit PROFIsafe Commissioning" in the
menu bar.
2. Press the "Quit PROFIsafe operation mode or commissioning" button.
3. Press the "Close" button.

Other manuals for Siemens SITRANS P series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SITRANS P series and is the answer not in the manual?

Siemens SITRANS P series Specifications

General IconGeneral
CommunicationHART, PROFIBUS PA, FOUNDATION Fieldbus
Housing MaterialStainless steel, aluminum
Measurement RangeVaries by model (e.g., 0-10 bar, 0-100 bar, etc.)
Output4-20 mA, HART, PROFIBUS PA, Foundation Fieldbus
Operating Temperature-40°C to +85°C (-40°F to +185°F)
Ingress ProtectionIP66, IP67
Measured MediumLiquids, gases
Process Temperature-40°C to +125°C (-40°F to +257°F)
Ambient Temperature-40°C to +85°C (-40°F to +185°F)
ApprovalsATEX, IECEx, CSA
Process Connection1/2" NPT