Alarm, error, and system messages
10.2 Alarm list
SITRANS SL
174 Operating Instructions, 12/2010, A5E01132948-04
Symbol/Group Error
Code
Alarm (description)
Explanation
/Action
E57 EEPROM configuration missing
EEPROM configuration missing.
Contact Siemens support.
E58 Save settings failed
Save settings failed.
Contact Siemens support.
E09 Save settings active Configuration changed
E55 Analog output test mode The test mode for analog outputs is active
E56 Digital output test mode The test mode for digital outputs is active
E25 PROFIBUS configuration error
PROFIBUS configuration fault.
Contact PLC support.
E26 PROFIBUS connection error
PROFIBUS Connection Error.
Check that PROFIBUS cables are properly connected.
E27 PROFIBUS master switch off
PROFIBUS Master Switch Off.
Contact PLC support. Check the PROFIBUS master
switch.
E28 PROFIBUS internal error
PROFIBUS internal error.
Contact Siemens support.
E61 LUI unlocked
The LUI is unlocked and has no password protection.
Lock the LUI.
E60 Watchdog reset triggered
The analyzer has been reset due to a triggered watchdog
event.
This alarm is automatically cleared at the screen after
restart of the system. The alarm will still be visible in the
logbook.