Safety Software Related Safety Information
0.0
04/2012
0.0
A.4-19
0.0
syngo Body Perfusion CT
A.4
syngo Body Perfusion CT provides images of blood flow, blood
volume and time to peak from one set of dynamic CT images.
A.4
Segmentation
A.4
A.4
A.4
A.4
CAUTION
A.4
Motion artifacts on images!
A.4
Incorrect automatic segmentation.
A.4
After loading of images check images for motion artifacts.
Also check results of segmentation. If not applicable, use
Reset button.
Remove image(s) with motion artifacts.
Then repeat evaluation.
A.4
CAUTION
A.4
Evaluation settings are not adapted!
A.4
Wrong diagnosis.
A.4
Check that the thresholds for CBF and CBV are the intended
ones.