Switching the System On and Off Operation
D.1-24 SOMATOM Spirit
0.0
Power failure
D.1
The uninterruptible power supply (UPS) takes over the power
supply of your computer system temporarily during a power fail-
ure.
D.1
However, during a power failure, the power supply of the mea-
suring system is not guaranteed.
D.1
The UPS outputs an audible alarm if the input voltage drops.
D.1
Warning
D.1
A message box is displayed when the public power supply fails.
As long as the UPS takes over the power supply of your com-
puter system, you have the option of completing current
actions.
D.1
If the power failure is remedied during this time, you can con-
tinue with normal operation after resetting the system.
D.1
Shutdown
D.1
When the backup time (3 min) of the UPS has elapsed, shut-
down is initiated.
D.1
Restart
D.1
As soon as the normal power supply is reactivated you can
restart your system.