0.0
0.0
Camera Test Quality Assurance
F.5-4 SOMATOM Spirit
0.0
Evaluating the test
F.5
Compare the test results with the reference values.
Measure the fields marked in the first test image with the
densitometer X-RITE 331.
Enter the measured values in the constancy test report.
fB
.5
F
Test is out of tolerance.
F. 5
Contact Siemens Customer Service.
F. 5
Example
F. 5
Here is an example of test image 1 with measurement fields
marked by the technologist:
F. 5
F. 5