Operation Switching the System On and Off
0.0
04/2012
0.0
D.1-17
0.0
Restart after System Emergency Stop
D.1
In case the system has been switched off by pressing the emer-
gency key in the examination room (for example, by accident),
a controlled restart has to be done.
D.1
B
D.1
Country-specific regulations must be observed.
D.1
Starting the system
D.1
All causes of danger have been found and remedied.
Wait for approximately 1 minute after power switch off.
Release the EMERGENCY OFF key provided by the cus-
tomer.
Switch the system on again.
Perform the usual checkup procedure to calibrate the sys-
tem.