Mechanical Safety Safety
A.3-4 SOMATOM Spirit
0.0
Explosion protection
A.3
The system and accessories are not suitable for operation or
storage in areas with explosive gases. The same applies to the
accessories.
A.3
Electromagnetic compatibil-
ity
A.3
The system complies with the EMC stipulations.
A.3
Certain items of equipment, for example, radio telephones
(“mobiles”) exceed the limit values of the EMC stipulations. In
extreme cases, use of such items can cause interference.
A.3
✧ Do not use a radio telephone or similar equipment in the
vicinity of the system.
Protection measures
A.3
The following protection measures must be followed when
installing the system.
A.3
Power connection
A.3
The power supply must be provided to all products operated as
part of an X-ray system through a fixed wiring connection and a
multipole interrupting device provided by the customer. The
equipment must be installed according to specification
DIN VDE 0100, Part 710 or must meet the respective national
regulations.