Safety Mechanical Safety
0.0
04/2012
0.0
A.3-5
0.0
Fire extinguishers
A.3
CO
2
fire extinguishers must be located in easily accessible and
visible locations.
A.3
Site on/off switch
A.3
In compliance with the MDD (Medical Device Directive) an on/
off switch must be installed on site by the customer.
A.3
The on/off state must be visible.
A.3
Radiation warning lamps
A.3
Radiation warning lamps must be installed on all doors of the
examination room. They must be visible from all areas where
radiation can be released.
A.3
Radiation warning acoustic
signal
A.3
A warning signal sounds when radiation is being generated.
A.3