CAUTION
A.4
If syngo receives images but the study or series UID is already
available in the local database, the data will be assigned to this
locally existing patient. This happens regardless of the patient
identification (name, date of birth, gender and patient ID) of the
received images. Changes of patient information on other sys-
tems will not automatically change the patient information in
syngo!
A.4
Data retrieved from PACS or other nodes seems to be mis-
aligned or lost, but is stored elsewhere in the database.
A.4
✧ Imported data will always be appended to the already exist-
ing patient on the local system, based on matching DICOM
UIDs. If you imported data, but cannot find them on the
local system under a certain name, use other criteria from
the study or series to search for the imported data, or filter
the data. To completely avoid these inconsistencies, always
delete the local patient data after you have successfully
archived them.