Switching the System On and Off Operation
D.1-10 SOMATOM Spirit
0.0
D.1
✧ Click on Checkup.
The system checkup is started.
D.1
You are prompted to press START.
D.1
D.1
✧ Press the Start key on the control box.
Scanning is initiated.
D.1
In the dialog box, you then read “Automatic checkup running”.
D.1
After a successful checkup, you read “Automatic checkup com-
pleted”.
D.1
The Checkup dialog box is closed automatically. The user inter-
face with the task cards is displayed.
D.1
fB
D.1
An error message is displayed.
D.1
An error occurred during checkup.
D.1
D.1
✧ Click on Retry in the dialog box displayed.
Checkup is started again.
D.1
D.1
fB
D.1
The message “Scan system should not be used” is displayed.
D.1
The checkup failed several times. The system is not ready for
examination.
D.1
✧ Call the system administrator and/or Siemens Customer
Service.