Operation Switching the System On and Off
0.0
04/2012
0.0
D.1-15
0.0
D.1
✧ Click on Yes.
All applications will be closed. The system will continue to shut
down.
D.1
or
D.1
D.1
✧ Click on No.
You return to the user interface.
D.1
✧ Exit all applications on the user interface.
Switching off the system
D.1
✓ If no errors occur during shutdown, the computer will be
switched off.
D.1
✧ Press the O pushbutton at the UPS at least 3 seconds.
The console will be switched off.
D.1
B
D.1
Unless absolutely necessary, do not switch off the system at the
main switch provided by the customer or the EMERGENCY OFF
switch. Otherwise you will have to calibrate the system several
times when you restart it. (→ Page D.1-8 System checkup and
calibration)
D.1