Switching the System On and Off Operation
D.1-20 SOMATOM Spirit
0.0
Restart
D.1
When the system is ready to be restarted after shutdown, a mes-
sage is displayed.
D.1
i
D.1
If the temperature in the gantry rises above a maximum value,
for example, as the result of a defective cooling system, the gan-
try is automatically switched off. A message appears stating
that the gantry is no longer ready for operation.
D.1
When it has cooled down, the gantry automatically switches on
again. A message stating this fact appears.
D.1
Resolving the temperature
warning
D.1
Temperature warning dialog box is displayed.
Countdown has started.
Finish the current examination and check the cooling sys-
tem.
Confirm with OK in the dialog box displayed.
The dialog box is closed.
D.1
i
D.1
A text is displayed in the status bar during countdown.
D.1