Switching the System On and Off Operation
D.1-22 SOMATOM Spirit
0.0
Warning
D.1
If the temperature of the computer system rises above a certain
threshold value, a dialog box indicates this to you.
D.1
Recovery
D.1
If the temperature of the system falls below the threshold value,
a message is displayed. You can continue operation.
D.1
If the temperature continues to rise and reaches a critical value,
countdown is initiated.
D.1
Countdown
D.1
As long as the countdown is running, you have the option of
completing current actions and storing your data. A dialog box
is displayed.
D.1
If the temperature falls below the threshold value during count-
down, you can continue operation.
D.1
Shutdown
D.1
If the countdown has finished and the temperature is still above
the critical value, the system is shut down. A message is dis-
played in the status bar.
D.1
Restart
D.1
When the computer system has cooled down, you can restart
the system.