Accessories for Patient Positioning Operation
D.3-8 SOMATOM Spirit
0.0
Warning label: Insert pads
D.3
D.3
Use of insert pads!
D.3
Increased risk of the head of the patient hitting the gantry.
D.3
Always watch the patient while the table is moving - espe-
cially if the gantry is tilted.
Warning label: Maximum
load
D.3
D.3
Overloading the head holder / receptacle of the patient table!
D.3
Damage to the head holder.
D.3
The maximum permissible weight must not exceed 180 N,
this corresponds to approximate 18 kg (40 lbs).
Using the CFK head holder
D.3
Use this accessory for positioning the head for cranial
examinations, for examinations in the neck region and for
examinations of the cervical vertebrae.