General Safety Information Safety
A.1-8 SOMATOM Spirit
0.0
System movements
A.1
When the gantry or the patient table is moved, the patient or
other people could be injured.
A.1
Make sure that the patient and other people are not in dan-
ger before you execute system movements or scans.
Ensure that no objects can obstruct movement.
Stopping system move-
ments and radiation
A.1
All system movements and radiation can be interrupted in case
of an emergency.
A.1
Make sure that you know where the STOP key is located.
Always observe the patient directly. Press the nearest STOP
key in case of danger.
B
.1
A
You will find STOP keys on the gantry and on the control box.
(→ Page A.2-14 Terminating system movements and radia-
tion)
A.1
Releasing the patient
A.1
After you have pressed a STOP or EMERGENCY OFF key or if a
power failure has occurred, no motorized system movements
are possible. However, you can pull the patient out of the open-
ing of the gantry manually.
A.1
Press the lever under the handle of the table top and pull.
(→ Page C.2-8 Moving the patient table manually)
Release the table top to retract the table top with the
patient.