EasyManua.ls Logo

Siemens SOMATOM Spirit - Page 392

Siemens SOMATOM Spirit
484 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
0.0
0.0
Constancy Test Quality Assurance
F.3-8 SOMATOM Spirit
0.0
Performing the lightmarker
test (z-position)
F.3
With the quality measurement, you determine the deviation of
inner light marker from the current slice plane.
F. 3
i
.3
F
After successful completion of the phantom position check the
Lightmarker test is started automatically without pressing the
Start key.
F. 3
The set of phantoms is positioned in such a way that the
inner light marker points on the reference marking on the
slice thickness phantom. (→ Page F.1-1 General Informa-
tion)
The Lightmarker test mode is loaded.
You are prompted to press Start.