Constancy Test Quality Assurance
F.3-32 SOMATOM Spirit
0.0
Performing the CTDI Air test
F.3
With this test, you measure the dose in the system axis.
F. 3
i
.3
F
After successful completion of the table position test the CTDI
Air test is started automatically without pressing the Start key.
F. 3
✓ The ionization chamber is centered in the axial direction in
the system axis. The slice plane has to run through the cen-
ter of the chamber. An appropriate dosimeter is connected
to the ionization chamber.
✓ The CTDI Air test mode is loaded.
✓ You are prompted to press Start.
F.3
✧ Press the Start key on the control box.
Scanning is started.
F. 3
F. 3
✧ Read off the value on the dosimeter and enter it in the pro-
tocol.
✧ Continue with the next kV values.