Constancy Test Quality Assurance
F.3-34 SOMATOM Spirit
0.0
Exiting the constancy test
F.3
After the final test has been completed, you terminate the
monthly constancy test and return to the Home Menu dialog
window.
F. 3
✓ In the status and error message area, you are prompted to
press Done.
F. 3
F.3
✧ Click Done.
The Constancy test is terminated.
F. 3
B
.3
F
In Reference mode, click the Save button first before terminat-
ing the constancy test with Done. Without Save the old refer-
ence values will be retained.
F. 3
F.3
✧ Click Home.
You return to the Home window.
F. 3
i
.3
F
You can now view the measurement report with the results of
the actual and previous tests. (→ Page F.3-35 Output of
results)
F. 3
✧ Close the Local Service window.
You return to the syngo user interface.
F. 3
✧ Confirm closing of the service session with OK in the dialog
displayed.