Software Related Safety Information Safety
A.4-4 SOMATOM Spirit
0.0
A.4
A.4
A.4
CAUTION
A.4
mA proposal of CARE Dose not adequate due to wrong attenua-
tion data from topogram!
A.4
X-ray not or only partially usable.
A.4
✧ Check plausibility of the automatically selected scan param-
eters before scanning.
A.4
CAUTION
A.4
RTP board is an additional absorber in the scan field!
A.4
Reduced low contrast resolution at soft tissue tumors.
A.4
✧ Consider reduced low contrast resolution when identifying
soft tissue tumors.