Maintenance
6-3
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
6.1.2 Fault Messages
Fault messages are references to certain changes in the analyzer
which influence its ability to measure correctly.
If the analyzer is in measuring mode, the occurrence of a fault is
indicatedbyan”F” at the right edge of the display.
Fault messages are logged and can be called in input mode using the
menu path ”Analyzer status - Status - Logbook/faults” (see Section
5.7.1.1). The corresponding message texts are stored as alphanumeric
text in the logbook. You can delete the messages by pressing the
ENTER key. However, they appear again if the cause has not been
eliminated.
Table 6-1 provides a summary of fault messages, their causes, and
measures to eliminate the faults. If a remedy is not listed, you must
contact the servicing department when this fault message occurs.
Table 6-1 Fault messages
Fault message
Possible causes Remedy
Meas. value channel 1 beyond
tolerance
Meas.value display: *****
Analyzer section of first
component is faulty
Meas. value channel 2 beyond
tolerance
Meas.value display: *****
Analyzer section of second
component is faulty
Meas. value channel 3 beyond
tolerance
Meas.value display: *****
Analyzer section of third
component is faulty
Meas. value O
2
beyond tolerance
Meas. value display: *****
O
2
sensor faulty or old Replace the O
2
sensor as
described in Section 6.2.1.
Motherboard faulty
Mains voltage beyond tolerance Mains voltage varies Correct such that the mains
voltage is within the tolerance
values permissible for the
analyzer.
Power supply unit faulty
Temp. of analyzer beyond
tolerance
Ambient temperature too high or
too low
Provide sufficient ventilation or air
conditioning.
Temperature sensor of a detector
faulty
Pressure of amb. air beyond
tolerance
Pressure sensor faulty
Mains voltage beyond
tolerance
press ENTER to clear
next message with - >