Operation
5-10
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
5.5.2 Key Operations Step-by -step
This section describes operation of the analyzer with the keys using an
example. The analyzer is in measuring mode (see Section 5.4).
S Change from measuring mode to input mode by pressing the
ENTER key. You first access the main menu. A cursor flashes on
the character ”A
” at the left edge of the first line.
S You can set the cursor to the start of each line using the ↑ and ↓
keys. Cursor movements are cyclic, i.e. if you move above the top
edge of the display, the cursor appears again in the bottom line, and
vice versa.
S Call the respective menu item by pressing the ENTER key.
The cursor is located at ”P
” when you have pressed the ↓ key twice.
S Now call the submenu ”Parameters” by pressing the ENTER key.
The adjacent display appears in which you will be requested to enter
the code number for code level 1.
S You can change the value of the code digit to which the cursor is
pointing using the ↑ and ↓ keys.
S Move to the next position of the code number using the → key. This
function is also cyclic, and the cursor appears at the first position
again when you move it beyond the last position
S Close the code input by pressing the ENTER key.
S Press the ENTER key again to call the submenu ”Measuring
ranges”.
S Now select components 1 to 4 for which the subsequently set
ranges are to apply. Up to four components can be present.
403.8 vpm CO |
1875 vpm NO |
936.4 vpm SO
2
|
4.2 % O
2
|
Analyzer status
Calibration
Parameters
Configuration
Analyzer status
Calibration
P
arameters
Configuration
Level 1 required
Please enter code
: 0
00
Level 1 required
Please enter code
:11
1
Measuring ranges
Limit values
Time constants
Pump/LCD contrast
Choose component
:NO 1